3. Installation
3.1.1 Gas sensor types for the TX-KFP (Cont’d)
3.1.2 Mounting of the gas sensor types BP and MP
The TX-KFP unit comes pre-assembled with items 1, 2 and 3.
To install the BP- or MP-type gas sensor into the TX-KFP body, proceed as follows.
For gas sensors with TG no.’s: 170/270/880/
1560 and 1750 the “BIAS voltage control” is
activated through a 1.5 Volts battery device,
see fig. 3.1.2.
Make sure that in case you have to reconnect
the device, the battery’s voltage has not
dropped under the required voltage level of
1.2 V.
- Remove the “BIAS voltage control” provision,
see section 3.1.1.
- Push the EP-adapter (5) gently upwards into the
KF body (1) through the retaining ring (4)
until it snaps into the sensor socket (2).
- Plug the BP- or MP-type sensor (6) into the EP-
adapter.
- Slide the fastening tool (7) over the sensor until
the tool projections snap into the notches of the
retaining ring (4). Turn anti-clockwise to secure
the sensor with the retaining ring.
Fig. 3.1.2