the measurement may be lowered, resulting in poor breathing. To prevent this,
connect all component is securely and check the connection according to standard
clinical procedures to ensure that there are no leaks.
WARNING
Industrial safety: Carefully dispose of used sampling tubes and T-connectors as they
may cause infection. There is a risk of infection. Dispose of all equipment in
accordance with local regulations.
Optimize reaction time by minimizing dead space and keeping sample collection
tubes as short as possible. Long sampling tubes can lead to poor accuracy and slow
response times for sidestream measurement techniques.
Do not place the airway adapter between the suction catheter and the endotracheal
tube when using the sample collection line as a closed suction device for tuberous
patients. This is to ensure that the airway adapter does not interfere with the function
of the suction catheter.