Rev. 2.2
59
is the silence period of the alarm, the alarm silence is canceled and the alarm sound is generated
again.
Alarm_Paused:
It transitions to a paused alarm for a user-defined time period, suspending visual and audible alarms.
A banner with the message Alarm Pause and a countdown timer are displayed on the screen. In
this state, if the alarm continues to occur even if the user presses the alarm mode key again or the
timeout period has elapsed during the alarm pause period, the alarm pause is canceled and the
alarm display and alarm sound are generated again.
Alarm_Off:
Stop visual and audible alarms. A banner with the message Alarm Off is displayed on the screen.
All alarms remain stopped until the user switches to another alarm mode.
Audio_Off:
Stop an Audible Alarm. A banner with the message Audio Off is displayed on the screen. The
monitor remains silent until the user switches to another alarm mode.
Alarm Control:
Various alarm functions, such as alarm hold, validity and alarm limit indicators, can only be
configured in the alarm control menu, accessible only through the password protected unit manager
menu.
Nurse Call:
If the monitor is sounding an alarm, the Nurse Call system is signaling.
When an audible alarm is silenced (Audio Pause or Audio Off) at the bedside unit, the Nurse Call
system will not alarm.
Your system administrator can change the alarm priority level for the Nurse Call signal.
Summary of Contents for BM1
Page 14: ...Rev 2 2 14 CAUTION Infectious devices and parts must be sanitized and cleaned before disposal ...
Page 19: ...Rev 2 2 19 Basic Unit Front View Alarm lamp Battery operation indicator AC status indicator ...
Page 22: ...Rev 2 2 22 Back Side View Speaker Hole IV Pole Mount Battery Cover ...
Page 55: ...Rev 2 2 55 ...
Page 62: ...Rev 2 2 62 Alarm Review ...
Page 89: ...Rev 2 2 89 NIBP Connector and Measurement Cable NIBP Connector Adult Cuff ...
Page 139: ...Rev 2 2 139 Symbols and degree s greater than less than minus number percent plus or minus ...