The end-tidal CO
2
value is the highest
peak CO
2
value of all end of
expirations (end of breaths) over the
selected time period. If less than two
breaths exist in the selected time
period, the value will be the
maximum ETCO
2
value for the last
two breaths.
20SEC
B-5-2.
BALANCE GAS
This setup mode to setup the gas in
the measurement.
the type of gas that is mixed with the
breathing gas measuring
ROOM AIR
N2O
HELIUM
B-5-3.
SLEEP MODE
Sleep mode is used to save power
when the host monitor is in standby
mode. There are two sleep modes
available for the Capnostat. Using
Sleep Mode 1 maintains the heaters
so the Capnostat is able to run
immediately after exiting the sleep
mode. Mode 2 will require the
Capnostat to go through its warm up
sequence when exiting this mode
and a delay will be introduced until
the system has stabilized.
NORMAL MODE
TURNOFF MODE
POWER SAVING
B-5-4.
BARO. PRESSURE
This setting is used to set current
Barometric Pressure.
400~850mmHg
(Default 760mmHg)
B-5-5.
GAS TEMPERATURE
This setting is used to set
temperature of the gas mixture. This
setting is useful when bench testing
using static gasses where the
temperature is often room
temperature or below.
0~50°C
(Default 35.0°C)
B-5-6.
O2 COMPENSATION
Use this setting to correct for the
compensation of the gas mixture
administered to the patient.
0~100
B-5-7.
ANESTHETIC AGENT
Anesthetic agent is ignored when the 0.0~20.0
Summary of Contents for BM1
Page 14: ...Rev 2 2 14 CAUTION Infectious devices and parts must be sanitized and cleaned before disposal ...
Page 19: ...Rev 2 2 19 Basic Unit Front View Alarm lamp Battery operation indicator AC status indicator ...
Page 22: ...Rev 2 2 22 Back Side View Speaker Hole IV Pole Mount Battery Cover ...
Page 55: ...Rev 2 2 55 ...
Page 62: ...Rev 2 2 62 Alarm Review ...
Page 89: ...Rev 2 2 89 NIBP Connector and Measurement Cable NIBP Connector Adult Cuff ...
Page 139: ...Rev 2 2 139 Symbols and degree s greater than less than minus number percent plus or minus ...