Message
Status
CHECK PROBE
Reusable finger probe is off the patient. Check the probe. The factory
default for this alarm is MESSAGE ALARM.
PULSE SEARCH
Detection by the monitor of a repeatable pulse has ceased. Check
the patient and the probe site.
POOR SIGNAL
The SpO2 signal is too low. No SpO2 data is displayed. This can be
due to a low patient pulse, patient motion, or some other
interference. Check the patient and the probe.
LOST SIGNAL
SPO2 data continues to be displayed, but the quality of the signal is
questionable. Check the patient and the probe.
ARTIFACT
It indicates that something happened to the pulses; determine if the
artifact to be abnormal and irregular.