Rev. 2.2
129
Manufacturer’s Declaration - Electromagnetic Emission
The BM1 system is intended for use in the electromagnetic environment specified below. The
customer or the user of BM1 system should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The BM1 system uses RF energy only for its internal
function. Therefore. Its RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment.
RF emissions CISPR 11
Class A
The BM1 system is suitable for use in all establishm
ents other than domestic and those directly conne
cted to the public low-voltage power supplies buil
dings used for domestic purposes.
Harmonics emission IEC
61000-3-2
A
Voltage fluctuation
IEC 61000-3-3
Complies
Manufacturer’s Declaration - Electromagnetic Immunity
Recommended Separation Distances Between Portable and Mobile RF Communications Equipment and
the BM1 system
The BM1 system is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The user of the BM1 system can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the BM1 system as recommended below, according to the maximum output power
of the communications equipment.
Rated maximum output
power (W) of transmitter
Separation distance (m) according to frequency of transmitter
150kHz to 80MHz
80MHz to 800MHz
800MHz to 2.5GHz
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.17
1.17
2.33
10
3.70
3.70
7.37
100
11.70
11.70
23.30
Summary of Contents for BM1
Page 14: ...Rev 2 2 14 CAUTION Infectious devices and parts must be sanitized and cleaned before disposal ...
Page 19: ...Rev 2 2 19 Basic Unit Front View Alarm lamp Battery operation indicator AC status indicator ...
Page 22: ...Rev 2 2 22 Back Side View Speaker Hole IV Pole Mount Battery Cover ...
Page 55: ...Rev 2 2 55 ...
Page 62: ...Rev 2 2 62 Alarm Review ...
Page 89: ...Rev 2 2 89 NIBP Connector and Measurement Cable NIBP Connector Adult Cuff ...
Page 139: ...Rev 2 2 139 Symbols and degree s greater than less than minus number percent plus or minus ...