MKF+ MKFT (E3.2) 07/2017
page 38/115
6.3
Menu settings in the “User-settings” menu
User-settings
Instrument data
Contrast
Displ. Power down
Contin. operation
User Level
Safety control. Act + 100.8 °C
Safety control. Set. + 110.0 °C
29
Instrument data
•
Instrument Name
Enter an individual name of the chamber.
•
Address
Enter a controller address (1 to 30) for operation with the communication software APT-COM™.
All other entries are relevant only for service purposes.
Contrast:
No function.
Displ. power down
•
Switch off event
Do not change the entry “Wait. Period”.
•
Waiting period
You can enter a delay time after which the display, following manual activation, will automatically be
turned off again, on condition that the moment is outside the operation time defined in menu ”Contin.
operation”.
Contin. operation
Enter an operation time to determine the period of display activity. Outside the defined period, the display
is automatically turned off. Pressing down any key will reactivate the display. After the time set in menu
“Displ. power down“, the display will turn off again as far as the actual time is not within the operation time
fixed in menu “Cont. operation“.
User Level
Toggle here to the display menu “User Level” (chap. 6.4) by entering a password. Factory default setting
for this password is +00001. You can change the password (“user code”) in the “User Level” menu.
Safety control.Set
The setting of the tolerance limit of the safety controller (chap. 12.2) is displayed. You cannot change it in
this view.
Safety control.Act
The measuring result of the safety controller is displayed. The safety controller compares the value
measured by a second independent temperature sensor to the entered tolerance limit.
Summary of Contents for MKF 115
Page 97: ...MKF MKFT E3 2 07 2017 page 97 115 20 10 Dimensions MKF 115 MKFT 115...
Page 98: ...MKF MKFT E3 2 07 2017 page 98 115 20 11 Dimensions MKF 240...
Page 99: ...MKF MKFT E3 2 07 2017 page 99 115 20 12 Dimensions MKFT 240...
Page 100: ...MKF MKFT E3 2 07 2017 page 100 115 20 13 Dimensions MKF 720 MKFT 720...
Page 102: ...MKF MKFT E3 2 07 2017 page 102 115...
Page 103: ...MKF MKFT E3 2 07 2017 page 103 115...
Page 104: ...MKF MKFT E3 2 07 2017 page 104 115 21 2 EU Declaration of Conformity for MKFT...
Page 105: ...MKF MKFT E3 2 07 2017 page 105 115...
Page 106: ...MKF MKFT E3 2 07 2017 page 106 115...
Page 108: ...MKF MKFT E3 2 07 2017 page 108 115...
Page 109: ...MKF MKFT E3 2 07 2017 page 109 115 22 Product registration...