MKF+ MKFT (E3.2) 07/2017
page 64/115
Checking and setting safety controller set-point type and safety controller set-point:
Unlock the keyboard locking (option, chap. 16.5).
Grundeinstellung
User-settings
Instrument data
Contrast
Displ. Power down
Contin. operation
User Level
Safety control. Act + 100,8 °C
Safety control. Set. + 105.0 °C
29
User level code no.
1
(factory setting)
In the menu “User Level” select the submenu “Safety controller”.
•
Select the safety controller set-point type “Limit” or “Offset” in the field “Setpoint type”
•
Enter the value for “Limit” or “Offset” in the field “Setpoint”.
Lock afterwards the keyboard locking (option, chap. 16.5).
For temperature disturbances see alarm indications, chap. 13.
User Level
Date
and time
Summer time
Language
Temperature unit
Buzzer
Safety controller
User-code No.
English
°C
1
Active
Safety Controller
Setpoint type
Setpoint
+ 110.0
Limit
Sollwert Art
Grenzwert
Offset
Setpoint
+0110.0
Sollwert Art
Grenzwert
Offset
Setpoint type
Limit
Offset
4 x
Summary of Contents for MKF 115
Page 97: ...MKF MKFT E3 2 07 2017 page 97 115 20 10 Dimensions MKF 115 MKFT 115...
Page 98: ...MKF MKFT E3 2 07 2017 page 98 115 20 11 Dimensions MKF 240...
Page 99: ...MKF MKFT E3 2 07 2017 page 99 115 20 12 Dimensions MKFT 240...
Page 100: ...MKF MKFT E3 2 07 2017 page 100 115 20 13 Dimensions MKF 720 MKFT 720...
Page 102: ...MKF MKFT E3 2 07 2017 page 102 115...
Page 103: ...MKF MKFT E3 2 07 2017 page 103 115...
Page 104: ...MKF MKFT E3 2 07 2017 page 104 115 21 2 EU Declaration of Conformity for MKFT...
Page 105: ...MKF MKFT E3 2 07 2017 page 105 115...
Page 106: ...MKF MKFT E3 2 07 2017 page 106 115...
Page 108: ...MKF MKFT E3 2 07 2017 page 108 115...
Page 109: ...MKF MKFT E3 2 07 2017 page 109 115 22 Product registration...