Wastewater connection for humidifying system
Fasten the wastewater hose to the wastewater connection “OUT” (17) (Figure 8) on the rear of the cham-
ber (olive
∅
14 mm). Observe the following points:
•
You can use a part of the supplied tap water hose as a drainage hose. In case another hose is used, it
has to be permanently resistant against at least 95 °C /
203 °F
.
•
Mount the wastewater hose with a maximum positive inclination of 1 m and a maximum total length of
3 m.
•
Protect both ends of the drainage hose with two of the four supplied hose clamps.
Wastewater is collected in an internal can with a volume of approx. 0.5 liters. It is pumped off
only when required, thus there is no continuous wastewater flow.
Protect the wastewater supply at both sides with the supplied hose clamps.
4.2
Freshwater supply for humidifying system
Connect the wastewater pipe
before
connecting the chamber to a freshwater pipe or filling the
water can.
You can supply the chamber with freshwater via a water pipe or by manually filling the internal water can.
It is not necessary to switch between both possibilities. When connecting to a water pipe, the water can is
automatically filled.
Water intake temperature NOT below +5 °C /
41 °F
and not exceeding 40 °C /
104 °F
.
CAUTION
Calcification of the humidifying system.
Damage to the chamber.
Operate the chamber with deionized (demineralized) water only.
Types of suitable water quality
•
Deionized water from a water treatment installation already existing at the customer's site. Conductivi-
ty from 1 µS /cm up to a maximum of 20 µS/cm. (Water, which is in equilibrium with the CO
2
in the air,
and has a conductivity below 1 µS/cm (ultrapure water), may cause acid corrosion due to its low pH.)
•
Water treated by the optional water treatment system BINDER Pure Aqua Service (disposable sys-
tem). A reusable measuring equipment to assess the water quality is included (chap. 16.6).
BINDER GmbH is NOT responsible for the water quality at the user’s site.
Any problems and malfunctions that might arise following use of water of deviating quality is
excluded from liability by BINDER GmbH.
The warranty becomes void in the event of use of water of deviating quality.
Summary of Contents for MKF 115
Page 97: ...MKF MKFT E3 2 07 2017 page 97 115 20 10 Dimensions MKF 115 MKFT 115...
Page 98: ...MKF MKFT E3 2 07 2017 page 98 115 20 11 Dimensions MKF 240...
Page 99: ...MKF MKFT E3 2 07 2017 page 99 115 20 12 Dimensions MKFT 240...
Page 100: ...MKF MKFT E3 2 07 2017 page 100 115 20 13 Dimensions MKF 720 MKFT 720...
Page 102: ...MKF MKFT E3 2 07 2017 page 102 115...
Page 103: ...MKF MKFT E3 2 07 2017 page 103 115...
Page 104: ...MKF MKFT E3 2 07 2017 page 104 115 21 2 EU Declaration of Conformity for MKFT...
Page 105: ...MKF MKFT E3 2 07 2017 page 105 115...
Page 106: ...MKF MKFT E3 2 07 2017 page 106 115...
Page 108: ...MKF MKFT E3 2 07 2017 page 108 115...
Page 109: ...MKF MKFT E3 2 07 2017 page 109 115 22 Product registration...