In case of contamination of the interior by biologically or chemically hazardous material, there are two
possible procedures depending on the type of contamination and charging material:
(1) Spray the inner chamber with an appropriate disinfectant.
Before start-up, the chamber must be absolutely dry and ventilated, as explosive gases may form dur-
ing the decontamination process.
(2) If necessary, have strongly contaminated inner chamber parts removed by an engineer for cleaning, or
have them exchanged. Sterilize the inner chamber parts in a sterilizer or autoclave.
In case of eye contact, the disinfectant spray may cause eye damage due to chemical
burns. Follow the operating instructions and safety hints labeled on the bottle of the disin-
fectant spray.
Recommended precautions: To protect the eyes use sealed protective goggles.
VORSICHT
Eye contact.
Eye damage due to chemical burns.
∅
Do NOT empty into drains.
Wear protective goggles.
After using the disinfectant spray, allow the chamber to dry thoroughly, and aerate it sufficient-
ly.
17.3 Sending the chamber back to BINDER GmbH
If you return a BINDER product to us for repair or any other reason, we will only accept the product upon
presentation of an authorization number that has previously been issued to you. An
authorization num-
ber
(RMA number) will be issued after receiving your complaint either in writing or by telephone
prior
to
your sending the BINDER product back to us. The authorization number will be issued following receipt of
the information below:
•
BINDER product type and serial number
•
Date of purchase
•
Name and address of the dealer from which you bought the BINDER product
•
Exact description of the defect or fault
•
Complete address, contact person and availability of that person
•
Exact location of the BINDER product in your facility
•
A contamination clearance certificate (chap. 23) must be faxed in advance
The authorization number must be applied to the packaging in such a way that it can be easily recognized
or be recorded clearly in the delivery documents.
For security reasons we cannot accept a chamber delivery if it does not carry an authorization
number.
Return address:
BINDER GmbH
Abteilung Service
Gänsäcker 16
78502 Tuttlingen
Germany
Summary of Contents for MKF 115
Page 97: ...MKF MKFT E3 2 07 2017 page 97 115 20 10 Dimensions MKF 115 MKFT 115...
Page 98: ...MKF MKFT E3 2 07 2017 page 98 115 20 11 Dimensions MKF 240...
Page 99: ...MKF MKFT E3 2 07 2017 page 99 115 20 12 Dimensions MKFT 240...
Page 100: ...MKF MKFT E3 2 07 2017 page 100 115 20 13 Dimensions MKF 720 MKFT 720...
Page 102: ...MKF MKFT E3 2 07 2017 page 102 115...
Page 103: ...MKF MKFT E3 2 07 2017 page 103 115...
Page 104: ...MKF MKFT E3 2 07 2017 page 104 115 21 2 EU Declaration of Conformity for MKFT...
Page 105: ...MKF MKFT E3 2 07 2017 page 105 115...
Page 106: ...MKF MKFT E3 2 07 2017 page 106 115...
Page 108: ...MKF MKFT E3 2 07 2017 page 108 115...
Page 109: ...MKF MKFT E3 2 07 2017 page 109 115 22 Product registration...