The 2-channel program controller MB1 controls the following values inside the chamber:
•
Channel 1: Temperature in °C. Range without humidity: -40 °C /
-40°F
up to 180 °C /
356°F
,
in climatic operation: +10 °C /
50°F
up to +95 °C /
203°F
.
•
Channel 2: Relative humidity in % r.H. Range 10 % r.H. to 98 % r.H.
You can enter the desired set point values in Manual Mode or Program Mode (chap. 5.2) on the display
controller. For the control range of temperature and relative humidity, see chap. 14.
W
X
08:43:55 15.12.13
CONFIG
VIEW->
PGM
TEMP
40.0
36.8
°C
HUMID
65.0
58.6
%rH
HAND
Set point values Actual values
Figure 16: Normal display of the MB1 program controller in Manual mode
5.2
Operating modes
The 2-channel program controller MB1 operates in 3 modes:
Idle Mode
The controller is not functional, i.e., there is no heating or refrigeration and
no humidification. The fan is off.
Manual Mode
(Fixed value operation)
(HAND)
The controller operates as a fixed-point control, i.e., set-points for tempera-
ture and humidity can be defined, which are then maintained (chap. 8).
Program Mode (AUTO)
An entered temperature and humidity program is run (chap. 9).
The 2-channel program controller MB1 permits programming temperature and humidity cycles.
The controller offers 25 program memory positions with 100 program sections each. The total number of
program sections of all programs is limited to 500.
Programming can be done directly through the keypad of the controller or graphically through the soft-
ware APT-COM™ 3 DataControlSystem (option, chap. 16.1) specially developed by BINDER.
5.3
Behavior after power failure
After power returns, the chamber continues to function in the original operating mode it was in previously
before an actual power failure had occurred. In Manual Mode (HAND), the controller regulates tempera-
ture and humidity to the last entered set-points, while in Program Mode (AUTO) it regulates temperature
and humidity to their set-points that were reached during the program operation. The power failure is not-
ed in the event list (chap. 6.2) however, no error message is displayed indicating that a power failure has
taken place.
5.4
Behavior when opening the door
When you open the door, temperature control (heating and refrigeration) immediately stops (the com-
pressor continues running for 5 minutes without cooling). Humidification and dehumidification are off. The
fan is off.
Summary of Contents for MKF 115
Page 97: ...MKF MKFT E3 2 07 2017 page 97 115 20 10 Dimensions MKF 115 MKFT 115...
Page 98: ...MKF MKFT E3 2 07 2017 page 98 115 20 11 Dimensions MKF 240...
Page 99: ...MKF MKFT E3 2 07 2017 page 99 115 20 12 Dimensions MKFT 240...
Page 100: ...MKF MKFT E3 2 07 2017 page 100 115 20 13 Dimensions MKF 720 MKFT 720...
Page 102: ...MKF MKFT E3 2 07 2017 page 102 115...
Page 103: ...MKF MKFT E3 2 07 2017 page 103 115...
Page 104: ...MKF MKFT E3 2 07 2017 page 104 115 21 2 EU Declaration of Conformity for MKFT...
Page 105: ...MKF MKFT E3 2 07 2017 page 105 115...
Page 106: ...MKF MKFT E3 2 07 2017 page 106 115...
Page 108: ...MKF MKFT E3 2 07 2017 page 108 115...
Page 109: ...MKF MKFT E3 2 07 2017 page 109 115 22 Product registration...