12. Temperature safety devices
12.1 Over temperature protective device (class 1)
The chamber is equipped with an internal temperature safety device, class 1 acc. to DIN 12880:2007. It
serves to protect the chamber and prevents dangerous conditions caused by major defects.
If the actual temperature exceeds the nominal temperature by approx. 20 °C, the over temperature pro-
tective device permanently turns off the chamber. The user cannot restart the device again. The protec-
tive cut-off device is located internally. Only a service specialist can replace it. Therefore, please contact
an authorized service provider or BINDER service.
12.2 Safety controller (over-temperature safety device class 2)
The chamber is equipped with an over temperature safety device class 2 acc. to DIN 12880:2007. It is
designated as the “safety controller”. This second, electrically independent temperature controller takes
over at a selectable set-point in case of fault. It serves to protect the charging material against extremely
high temperatures.
With the option over-/under temperature safety device class 2 (chap. 12.3), the safety control-
ler must be set to maximum temperature.
The safety controller only activates after the set-point has been reached once.
The message “TEMPERATURE LIMIT” on the controller display indicates safety controller activity. The
safety controller controls the chamber to the entered safety controller set-point until the temperature in-
side the chamber returns below this temperature and until you then reset the alarm message by button
RESET.
Regularly check the safety controller setting for set-point type “Limit” or “Offset”
•
in Manual Mode according to the entered set-point temperature value
•
in Program Mode according to the highest temperature value of the selected temperature
program
Set the safety controller set-point by approx. 10 °C above the highest temperature set-point.
Safety controller set-point types:
Limit
Absolute maximum permitted temperature value.
Example:
Temperature set-point 100 °C/
212 °F
Limit value (safety controller set-point) set to 110 °C.
Offset
Maximum over temperature above the active temperature set point. The maximum tem-
perature changes internally and automatically with every set-point change it.
Example:
Temperature set-point 100 °C /
212 °F
Offset value (safety controller set-point) set to 10 °C.
Do NOT change the temperature unit from °C to °F.
Summary of Contents for MKF 115
Page 97: ...MKF MKFT E3 2 07 2017 page 97 115 20 10 Dimensions MKF 115 MKFT 115...
Page 98: ...MKF MKFT E3 2 07 2017 page 98 115 20 11 Dimensions MKF 240...
Page 99: ...MKF MKFT E3 2 07 2017 page 99 115 20 12 Dimensions MKFT 240...
Page 100: ...MKF MKFT E3 2 07 2017 page 100 115 20 13 Dimensions MKF 720 MKFT 720...
Page 102: ...MKF MKFT E3 2 07 2017 page 102 115...
Page 103: ...MKF MKFT E3 2 07 2017 page 103 115...
Page 104: ...MKF MKFT E3 2 07 2017 page 104 115 21 2 EU Declaration of Conformity for MKFT...
Page 105: ...MKF MKFT E3 2 07 2017 page 105 115...
Page 106: ...MKF MKFT E3 2 07 2017 page 106 115...
Page 108: ...MKF MKFT E3 2 07 2017 page 108 115...
Page 109: ...MKF MKFT E3 2 07 2017 page 109 115 22 Product registration...