C.2 Sensor Specifications
Sensor glucose
assay method
Amperometric electrochemical sensor
Sensor glucose
reading range
40 to 400 mg/dL
Sensor size
0.19 in height and 1.38 in diameter (5 mm
in height and 35 mm diameter)
Sensor weight
0.18 oz (5 grams)
Sensor power
source
One silver oxide battery
Sensor data
Up to 14 days
Sensor memory
8 hours (glucose readings stored every 15
minutes)
Sensor
transmission range
20 feet (six meters) unobstructed
Operating
temperature
50 °F to 113 °F (10.0 °C to 45.0 °C)
Sensor Applicator
and Sensor
Pack storage
temperature
36 °F to 82 °F (2.2 °C to 27.8 °C)
Operating and
storage relative
humidity
10-90%, non-condensing
Sensor water
resistance
IP27: Can withstand immersion into three
feet (one meter) of water for up to 30
minutes. Protected against insertion of
objects > 12 mm diameter
Operating and
storage altitude
-1,250 ft (-381 m) to 10,000 ft (3,048 m)
Radio Frequency
2.402 to 2.480 GHz BLE; GFSK; 0dBm
EIRP