Problem
What it may mean
Your Sensor is too hot (above its normal
operating range) to provide a glucose
reading.
What to do
Try moving to an area with an ambient temperature of 50°F to
113°F (10°C to 45°C). Wait 10 minutes for your Sensor to reach the
new temperature and begin sending glucose readings again.
Problem
What it may mean
Your Sensor is too cold (below its normal
operating range) to provide a glucose
reading.
What to do
Try moving to an area with an ambient temperature of 50°F to
113°F (10°C to 45°C). Wait 10 minutes for your Sensor to reach the
new temperature and begin sending glucose readings again.
Problem
What it may mean
Glucose data from
my Sensor did
not appear on my
App screen after I
scanned my Sensor
with my phone.
Your phone should only be used to scan
your Sensor when starting a new Sensor
and not beyond start-up. Do NOT scan
your Sensor with your phone to collect or
display current glucose data.
What to do
Always use your White Cap to scan your Sensor to collect and
display glucose data on your History screen.