Better Water LLC; rev. Jun 2016
Page 41 of 72
MediPac, Tank-Feed Operator Manual
SWITCHING TO TEMPORARY DI
*Reference the
VALVES LEGEND and OPERATION
section, previously detailed in this manual
for visual representation and valve numbers corresponding to the instructions below.
CONNECTING DI TANKS
1.
Turn the RO and Repress Pump OFF
…
- Turn the
RO’s
OPERATE-DISINFECT-OFF Switch
to
OFF
-
Turn Repress Pump off, by turning the
MediP
ac’s
REPRESSURIZATION PUMP Switch
to
OFF
2.
CLOSE
the
Repressurization Pump Inlet Valve (
bottom of the Storage Tank
).
3.
By-pass the Water Softener by
CLOSING
the Water Softener Inlet Valve (#4) and Water
Softener Outlet Valve (#5)
, and
OPENING
the
Water Softener By-Pass Valve
.
* This step is not required, but by-passing the softener will help lengthen the life of the DI resin.
4.
Remove the
Short Bowl or False Cartridge
from
the
Post-Treatment 5 Micron Post-DI Filter
and install the
10” Big Blue Housing
and
5
Micron-
10” Filter
.
5.
CLOSE
DI Water Inlet Valve (#19)
and
DI Water Outlet Valve (#20)
.
6.
CLOSE
the
Loop Return Valve
at the end of the loop at the Storage Tank.
7.
CLOSE
RO Inlet Valve (#14)
.
8.
OPEN
the
Drain Port on the Pre-Treatment 5 Micron Filter
to relieve system pressure.
9.
Disconnect the hose that goes between the
Pre-Treatment 5 Micron Filter
and the
RO
, at the
RO Inlet Connection
.
10.
Connect a
By-pass Hose
from that
RO Inlet Connection
to the
Post-RO and Storage Tank
Emergency By-pass Valve (#15)
.
RO Inlet Connection
By-pass Hose
Post-RO/Storage
Tank Emergency
By-pass Valve
(#15)