MediPac, Tank-Feed Operator Manual
DAILY START-UP
During normal operation, the RO would be set in
FLUSH MODE
at end-of-day, and therefore
should be in this mode when ready to start-up at the beginning of a new day.
1.
Verify that all valves are in the correct position for daily operation.
2.
Verify the
City Booster Pump
is
ON
:
-
For the Grundfos MQ Series… if the PUMP ON Light is not illuminated then press the
ON/OFF
Button
located on the pump.
- For the Walrus TQ-
800… verify that the POWER SWITCH located on the pump is
ON
.
3. Verify the Post-Treatment is ON:
On the MediPac Control Panel…
a.
Verify that the
REPRESSURIZATION PUMP Switch
is in the
ON
position.
b.
Verify that the
TANK ALARM Switch
is in the
ON
position.
4. Start the RO:
a.
Check the RO CONTROL PANEL:
-
All alarm lights should be
OFF
and
NO
audible alarm should be sounding.
- If any one or more of the alarm lights are
ON
and/or an audible alarm is sounding,
press
the
ALARM RESET Button
. All alarm lights should go out and audible
alarms should stop sounding.
b.
Verify that the
OPERATE-DISINFECT key switch
is in the
OPERATE
position.
If this switch is in the DISINFECT position, you MUST verify that the RO
does NOT contain any disinfecting solution BEFORE proceeding to the next step.
c.
Verify the
OPERATE-DISINFECT-OFF switch
is in the
OPERATE
position.
If the switch is in the
DISINFECT
or
OFF
position, see
step 2
and the warning above.
d.
If you must reset the switches listed above, the alarm lights will illuminate and
the audible alarm will sound, press
the
ALARM
RESET Button
to return the
alarms to normal.
WARNING
Disinfectants can cause serious injury or death to patients undergoing hemodialysis
treatment.