Better Water LLC; rev. Jun 2016
Page 48 of 72
MediPac, Tank-Feed Operator Manual
SYSTEM MAINTENANCE, Change Particulate Filter Cartridges
and Pyrogen Filters
Better Water LLC, recommends changing these filters as follows:
-
5 micron Big Blue Filters
: change monthly or
when the ΔP (
pressure drop
) reaches or
exceeds 15 psi while the RO is running
-
.03 micron Pyrogen Filters
: change every 6 months or
when the ΔP (
pressure drop
) reaches
or exceeds 15 psi while the RO is running
1.
Start with all valves, sample ports, vents, and drain ports closed.
2.
Open the drain port on the bottom of the housing to allow the water to drain completely from
the housing. Close the drain port when water flow stops.
3.
Depending on the filter housing either open it, or use a filter wrench to gain access to the filter.
4.
Remove the old filter(s), and wipe out the inside of the housing with a clean, damp cloth.
5
. Insert the new filter(s).
6.
Verify any o-rings are properly seated.
7.
Depending on the filter housing, either close it, or use a filter wrench to seal the cover, without
over-tightening.
8.
Slowly open the inlet valve to allow water to begin filling the housing.
9.
Partially open the sample port or vent to allow air to be purged from the device.
10.
When water flows from the sample port or vent, signaling that the air has been purged from
the device, close the sample port or vent.
11.
Slowly open the outlet valve fully to allow water to flow through the device and check for
leaks.
12.
The filter is now on-line and ready for use.
part#
SUCAOO00551
1
0” Big Blue
5 Micron Filter
part#
SUCAPE00575
20” .03 Ultra Filter
part#
SUMIOO00589
O-Ring for 2
0” Big
Blue Housing
part#
SUMIOO00587
O-Ring for 1
0” Big
Blue Housing