Better Water LLC; rev. Jun 2016
Page 50 of 72
MediPac, Tank-Feed Operator Manual
Changing Regenerate/Backwash Program Cycles:
The Fleck valve time settings for both filters and softeners have been provided for reference
below:
NOTE
The backwash and regeneration cycles are critical procedures in the pretreatment
process. There is a very balanced, structured, and programmed relationship between
the different pre-treatment component devices and the RO. Because of these, great
care must be taken when making any changes to the backwash and regeneration cycle
programs that have been setup by the manufacturer or authorized dealer. Before
considering or attempting to make any changes to these cycles by changing the pin
and hole settings on the regeneration cycle program wheel, technical support should
be consulted to advise and support any such changes.
WARNING:
There can be dire consequences to the quality of the water produced by the
pre-treatment system which can affect the RO and ultimately any patient undergoing
hemodialysis treatment if the backwash and regeneration cycles are programmed
incorrectly.