MediPac, Tank-Feed Operator Manual
Rebound Break
The Rebound Break is for 15-20 minutes. This
is highly recommended
as some residual
disinfectant could still be present in the membranes even if a negative result was obtained during
rinsing. After the rebound break time has been completed (15-20 min), restart the RO in normal
Operate mode again and recheck for Residual. If the residual is positive, continue to rinse until a
negative result is achieved. Once the residual is negative, turn off the RO and allow for another
Rebound Break of 15-20 minutes. Turn on the RO and re-check for residual. Repeat this
process as many times as necessary until a negative Residual is obtained after the rebound
break.
-
Place the RO in “Tank-Feed” mode, by opening the RO Control Box and turning the
Tank/Direct Switch
to
TANK
.
DISINFECT the POST-TREATMENT, STORAGE TANK, and
DISTRIBUTION LOOP
* Complete disinfect procedure of the RO first.
1.
Confirm that there is no need for purified water.
2.
Notify all necessary personnel that the system will be disinfecting.
3.
Fill the Storage Tank with the appropriate amount of RO water for the disinfect process.
4.
Turn
the RO and Repress Pump OFF…
- Turn the
RO’s
OPERATE-DISINFECT-OFF Switch
to
OFF
-
Turn Repress Pump off, by turning the
MediP
ac’s
REPRESSURIZATION
PUMP Switch
to
OFF
5.
If D.I. tanks are on-
line…
-
CLOSE
DI Inlet Valve
and
DI Outlet Valve
.
- Disconnect them from the system.
- Connect jumper hose(s) from DI Inlet Valve to DI Outlet Valve.
-
OPEN
DI Inlet Valve
and
DI Outlet Valve
.
6.
If a 5 micron filter is installed in the post-treatment filter housing (post-
DI)…
-
CLOSE
DI Outlet Valve
.
- Remove this filter and discard it.
- Install the false cartridge tube.
-
OPEN
DI Outlet Valve
.
7.
Remove the Storage Tank Lid and add appropriate amount of disinfectant to storage tank (
see
General Cleaning/Disinfecting Information
section in this manual
).
- Replace Storage Tank Lid.
8.
Start the Repressurization Pump
by…
- Set the MediP
ac’s
REPRESSURIZATION PUMP Switch
to
ON
- Press and hold the
PUMP START Button
until the repress pump starts
- Allow this to recirculate for 15-20 minutes.
- After 15-20 minutes, verify at each of the stations and at the
Loop
Return Sample Port
that there is disinfectant present, then let disinfectant
dwell
(circulate)
in the system for 30 minutes to 2 hours.