Chapter 9. Calibration and Advanced
72
MicroPolar 2 (++) LB 567
9.2
Adjusting the Calibration
A correction factor and an offset may be entered later to obtain
a subsequent adjustment of the calibration (fine tuning).
Below please find an example for an offset adjustment.
The display to the left appears if you push
RUN
.
The display reading is now compared with the analysis value
of the lab sample. The difference has to be entered as offset
with the correct algebraic sign.
Calculation:
Analysis value – Display = Offset
Eq. 9-1
Push
ESC
to return to the main menu.
SETUP
CALIBRATION
CALIBRATE CONC
1 | - | Live Display | 07.05 – 13:25
Concentration av.
65.50
%
Conc. av. Conc. act. 64.35%
ESC SAMPLE
..
▲▼
..
ZOOM
1 | - | LB 567 | 07.05 – 13:25
Live Display
Diagnostic
Setup
Access Level
User Mode
Language
English
STOP ▲ ▼ ►
1 | - | Setup | 07.05 – 13:25
Configuration
Calibration
Input / Output
Service
Product
Change password
⌂◄ ▲ ▼ ►
1 | - | Calibration | 07.05 – 13:25
System Adjust
Calibrate Conc
Advanced
⌂◄ ▲ ▼ ►
Summary of Contents for LB 567
Page 2: ......
Page 3: ......
Page 36: ......
Page 60: ......
Page 72: ...Chapter 7 Certificates 72 MicroPolar 2 LB 567 7 2 Frequency License...
Page 74: ...Chapter 8 Technical Drawings 74 MicroPolar 2 LB 567 8 2 Electrical Wiring Diagram...
Page 88: ...Chapter 8 Technical Drawings 88 MicroPolar 2 LB 567...
Page 89: ...Chapter 8 Technical Drawings MicroPolar 2 LB 567 89 8 4 2 Spiral antenna...
Page 94: ...Notes 94 MicroPolar 2 LB 567 Notes...
Page 95: ...Notes MicroPolar 2 LB 567 95 Notes...
Page 96: ......
Page 98: ......
Page 102: ......
Page 104: ......
Page 106: ......
Page 108: ......
Page 136: ......
Page 142: ......
Page 156: ......
Page 172: ......
Page 176: ......
Page 185: ...Notes MicroPolar 2 LB 567 89 Notes...
Page 186: ...Notes 90 MicroPolar 2 LB 567 Notes...