Chapter 4. Getting Started
42
MicroPolar 2 (++) LB 567
Figure 4-4:
Flow cell
connection,
version 1
Evaluation unit
Bottom side
M-Tx
M
-T
x
M-Rx
M
-R
x
R-Tx
R
-T
x
R-Rx
R
-R
x
Reference line
with N-connector
Flow cell
Bending radius
min. 100 mm
HF cable quad
(standard 2 m)
HF connections
max. 2 Nm
Flange
200 ... 250 Nm
M
-T
x
(
M
-R
x
)
M
-R
x
(
M
-T
x
)
R-Rx
R-Tx
Figure 4-5:
Flow cell
connection,
version 2
M-Tx
M-Rx
R-Tx
R-Rx
Evaluation unit
Bottom side
Reference cable
Flow cell
Bending radius
min. 100 mm
2 x sibnal cable
(standard 2 m)
HF connections
max. 2 Nm
Flange
200 ... 250 Nm
Hand tighten all screwed connections of the HF cable (2 Nm =
0.2 kg/m)! Before tightening, carefully screw on the cable by
hand.
Caution! Threaded joint jams easily.
Check occasionally if the screwed connection is still properly
tightened. If the installation is exposed to vibrations, the screwed
connection may come loose and this may result in inaccurate
measurements or corrosion of the connections.
As long as the cables are not connected, the coaxial sockets have
to be covered with plastic caps and the cable connectors have to
be protected against moisture and dirt.
Summary of Contents for LB 567
Page 2: ......
Page 3: ......
Page 36: ......
Page 60: ......
Page 72: ...Chapter 7 Certificates 72 MicroPolar 2 LB 567 7 2 Frequency License...
Page 74: ...Chapter 8 Technical Drawings 74 MicroPolar 2 LB 567 8 2 Electrical Wiring Diagram...
Page 88: ...Chapter 8 Technical Drawings 88 MicroPolar 2 LB 567...
Page 89: ...Chapter 8 Technical Drawings MicroPolar 2 LB 567 89 8 4 2 Spiral antenna...
Page 94: ...Notes 94 MicroPolar 2 LB 567 Notes...
Page 95: ...Notes MicroPolar 2 LB 567 95 Notes...
Page 96: ......
Page 98: ......
Page 102: ......
Page 104: ......
Page 106: ......
Page 108: ......
Page 136: ......
Page 142: ......
Page 156: ......
Page 172: ......
Page 176: ......
Page 185: ...Notes MicroPolar 2 LB 567 89 Notes...
Page 186: ...Notes 90 MicroPolar 2 LB 567 Notes...