33
www.bempu.com
4.4 Specifications
General Specifications
Intended User
Preterm Infants born under 36
weeks of GA (with
stimulation)
All Infants up to 1 year of age
(without stimulation)
Stimulus Type
Vibro-tactile
Physical Characteristics
Pulse Oximeter Probe length
1m (with 1m extension
cable)
Stimulator Boot length
2m
Device dimension
H: 10cm, W: 22cm, D: 7cm
Device weight
1 Kg approx
Material
Enclosure Plastic
ABS, UL94 V0 (FR grade)
Rubber Parts
Biocompatible Silicone as
per ISO 10993
Pulse Oximeter Specifications (NONIN OEM III Module)
Oxygen Saturation Range
(SpO
2
) 0-99%
Pulse Rate Range (PR)
18 to 250 beats per
minute (bpm)
SpO
2
Accuracy
70-100%
No Motion: ±3 digits
Motion: ±3 digits
Low Perfusion: ±4 digits
Pulse Rate Accuracy for
No Motion (18-300
Infants
BPM): ±3 digits
Motion (40-240 BPM): ±5
digits
Low Perfusion (40-240
BPM): ±3 digits
Summary of Contents for ApneBoot
Page 1: ...OPERATION MANUAL TM ...
Page 6: ...6 www bempu com ...
Page 10: ...10 www bempu com ...
Page 26: ...26 www bempu com ...
Page 42: ...42 www bempu com ...