96
BD FACSCanto II Flow Cytometer Reference Manual
Instrument Troubleshooting
Observation
Possible Causes
Recommended Solutions
Flow cell will not fill
Air in bubble filter
Run the Bubble Filter Purge and
De-gas Flow Cell command at
least once (you might need to run
it several times). See Purging the
Bubble Filter on page 78.
Fluidics cart power off
Turn on the power to the fluidics
cart by resetting the fluidics cart
circuit breaker switch (Figure 1-9
on page 28). Always use the
cytometer power button, located
on the left side of the cytometer, to
turn the system off and on.
No pressure in the plenum
Check air supply components (see
Figure 4-6 on page 87).
• Check air supply connections.
• Check air supply tubing for
kinks.
Sheath line disconnected
1
Check cubitainer-to-fluidics
cart, and fluidics cart–to-
cytometer connections. See
Reconnecting the Fluidics Cart
Tubing on page 86.
2
Check tubing for kinks.
BD FACSFlow cubitainer
empty
Replace the BD FACSFlow
cubitainer. See Changing a
Cubitainer on page 68.
Air in BD FACSFlow filter
(fluidics cart)
Purge air from the filter. See
Purging the Fluidics Filters on
page 57.
Air lock in BD FACSFlow
filter (on fluidics cart)
Summary of Contents for FACSCanto II
Page 4: ......
Page 12: ...xii BD FACSCanto II Flow Cytometer Reference Manual...
Page 44: ...44 BD FACSCanto II Flow Cytometer Reference Manual...
Page 104: ...104 BD FACSCanto II Flow Cytometer Reference Manual...
Page 124: ...124 BD FACSCanto II Flow Cytometer Reference Manual...
Page 138: ...138 BD FACSCanto II Flow Cytometer Reference Manual...
Page 144: ...144 BD FACSCanto II Flow Cytometer Reference Manual...
Page 152: ...152 BD FACSCanto II Flow Cytometer Reference Manual...