
38
BD FACSCanto II Flow Cytometer Reference Manual
2
If needed, remove the carousel from the Loader:
•
Slide the Loader doors open.
•
Pull out the Loader drawer.
•
Place your thumb on the center spindle, and press down with your
thumb while gently pulling up on the carousel handle with two fingers,
as shown.
3
Vortex the sample tubes and place them uncapped in the carousel(s)
according to the worklist.
For accurate results, match the tubes to those listed on the printed worklist
(BD FACSCanto clinical software) or the Carousel Assignment tab
(BD FACSDiva software). Print a copy of tube assignments and use the
printout as a guide when filling each carousel.
To prevent binding during loading, make sure label thickness per tube is
less than three labels. Flatten labels completely before placing a tube in the
carousel.
Summary of Contents for FACSCanto II
Page 4: ......
Page 12: ...xii BD FACSCanto II Flow Cytometer Reference Manual...
Page 44: ...44 BD FACSCanto II Flow Cytometer Reference Manual...
Page 104: ...104 BD FACSCanto II Flow Cytometer Reference Manual...
Page 124: ...124 BD FACSCanto II Flow Cytometer Reference Manual...
Page 138: ...138 BD FACSCanto II Flow Cytometer Reference Manual...
Page 144: ...144 BD FACSCanto II Flow Cytometer Reference Manual...
Page 152: ...152 BD FACSCanto II Flow Cytometer Reference Manual...