60
BD FACSCanto II Flow Cytometer Reference Manual
A completion message displays when the cleaning cycle finishes.
Figure 4-1
Long Clean completion message
4
Click OK.
5
Choose to shut down or continue.
•
To shut down, quit the software and turn off the power to the
cytometer.
•
To continue, select Cytometer (Instrument) > Fluidics Startup.
BD FACS shutdown solution can lyse cells. The Fluidics Startup
procedure removes BD FACS shutdown solution from the interior
reservoir and fluid lines and replaces it with BD FACSFlow solution.
!
Tip
If setup fails after a long clean, repeat Fluidics Startup until setup
passes.
Replacing the Air Filter
The BD FACSCanto II instrument has an air filter located in the side door.
Replace the filter every 6 months to ensure proper instrument performance.
If the completion message has not displayed after 90 minutes, verify there
are no error messages in the Status tab of the Instrument frame. If the
cleaning mode fails, see Fluidics Cart Troubleshooting on page 101.
BD FACSCanto clinical software
BD FACSDiva software