Chapter 5: Troubleshooting
101
Fluidics Cart Troubleshooting
Observation
Possible Causes
Recommended Solutions
Water leakage around
fluidics cart base
• Normal condensation
overflow from pressure
relief valve
• Extremely humid climate
1
Turn off the cytometer power.
2
Clean up the water.
3
Empty the condensation trap in
the fluidics cart daily. See
Figure 1-7 on page 26.
Bleeder valve on fluidics cart
filter open
1
Turn off the cytometer power.
2
Clean up the liquid.
3
Check and close all bleeder
valves. See Purging the Fluidics
Filters on page 57.
Broken fluid line
Contact BD Biosciences.
Fluidics cart will not
power on, cytometer on
Cart circuit breaker off
Switch on the cart circuit breaker.
Auxiliary air supply on, cart
not normally connected to
auxiliary air
Toggle the auxiliary air supply off.
When auxiliary air is on, the cart’s
own air pumps turn off.
Power cord from fluidics cart
to cytometer disconnected
Connect both ends of the cord.
Fuse blown
Replace the fluidics cart fuses. See
Replacing the Fluidics Cart Fuses
on page 92.
Summary of Contents for FACSCanto II
Page 4: ......
Page 12: ...xii BD FACSCanto II Flow Cytometer Reference Manual...
Page 44: ...44 BD FACSCanto II Flow Cytometer Reference Manual...
Page 104: ...104 BD FACSCanto II Flow Cytometer Reference Manual...
Page 124: ...124 BD FACSCanto II Flow Cytometer Reference Manual...
Page 138: ...138 BD FACSCanto II Flow Cytometer Reference Manual...
Page 144: ...144 BD FACSCanto II Flow Cytometer Reference Manual...
Page 152: ...152 BD FACSCanto II Flow Cytometer Reference Manual...