BD FACSCanto II Flow Cytometer Reference Manual
Replacing the Fluidics Cart Fuses
Power surges and other electrical events could cause a fuse to blow. Use the
following procedure to replace the fuses.
Materials
•
two replacement fuses (see Installation Kit on page 140 for ordering)
•
small screwdriver
Removing Old Fuses
1
Shut down the cytometer and turn off the instrument power.
2
Unplug the power cord from the wall outlet and the fluidics cart.
Removing the plug allows easier access to the fuse door.
3
Insert a small screwdriver into the slot and gently pry the fuse door open.
To protect against shock, always turn off the cytometer and unplug the
power cord before performing this procedure.
slot for screwdriver
empty socket,
plug removed
fuse door
circuit breaker