98
BD FACSCanto II Flow Cytometer Reference Manual
Cytometer on, no
response to software
commands
Bad keyboard or mouse
connection
Check keyboard or mouse
connections to computer. Refer to
the documentation that came with
your workstation.
Communication failure
(Ethernet error) between
computer and instrument
1
Turn off the computer and the
instrument.
2
Reseat the Ethernet cable,
located next to the power cord
on the right side of the flow
cytometer.
3
Turn on the instrument, then
the computer.
Software does not
connect to cytometer
BD FACSDiva software
running at same time as
BD FACSCanto clinical
software
Quit the software you are not
using or put the cytometer in
standby: Cytometer (Instrument)
> Standby.
Cytometer power is off
Turn power on.
Internal firmware error
Cycle power on cytometer.
Cytometer and fluidics
cart will not turn on
Power cord disconnected
from wall socket or cytometer
Reconnect the power cord to the
wall socket or the cytometer.
Instrument circuit breaker
tripped
Switch on the cytometer circuit
breaker, followed by the
instrument power (if off).
Blown fuse in cytometer
controller
Turn off the instrument main
power and call BD Biosciences.
Instrument Troubleshooting (continued)
Observation
Possible Causes
Recommended Solutions
Summary of Contents for FACSCanto II
Page 4: ......
Page 12: ...xii BD FACSCanto II Flow Cytometer Reference Manual...
Page 44: ...44 BD FACSCanto II Flow Cytometer Reference Manual...
Page 104: ...104 BD FACSCanto II Flow Cytometer Reference Manual...
Page 124: ...124 BD FACSCanto II Flow Cytometer Reference Manual...
Page 138: ...138 BD FACSCanto II Flow Cytometer Reference Manual...
Page 144: ...144 BD FACSCanto II Flow Cytometer Reference Manual...
Page 152: ...152 BD FACSCanto II Flow Cytometer Reference Manual...