Chapter 5: Troubleshooting
99
Tube does not load, or
sample tube does not fit
snugly on SIT
Improper tubes used
Use only uncapped 12 x 75-mm
BD Falcon polystyrene test tubes,
BD Trucount tubes, and BD FACS
7-color setup beads tubes. See
System Requirements on page 31.
Defective or cracked tube
• Transfer sample to new tube.
• Make sure you are using
appropriate tubes. See System
Requirements on page 31.
Bal seal improperly installed
or worn
Reinstall or replace the Bal seal.
See Replacing the Bal Seal on
page 79.
Liquid leakage around
cytometer base
Interior valve failure
1
Turn off the cytometer power.
2
Clean up the liquid, using
proper precautions.
3
Call BD Biosciences.
QC fails after Long
Clean
Residual BD FACSClean
solution in lines
Run Fluidics Startup to flush the
system with sheath fluid. Repeat
until QC passes.
Low laser power
indication
Flow cell access door open
Close the door completely.
Laser power output is below
requirement
Call BD Biosciences.
Instrument Troubleshooting (continued)
Observation
Possible Causes
Recommended Solutions
Summary of Contents for FACSCanto II
Page 4: ......
Page 12: ...xii BD FACSCanto II Flow Cytometer Reference Manual...
Page 44: ...44 BD FACSCanto II Flow Cytometer Reference Manual...
Page 104: ...104 BD FACSCanto II Flow Cytometer Reference Manual...
Page 124: ...124 BD FACSCanto II Flow Cytometer Reference Manual...
Page 138: ...138 BD FACSCanto II Flow Cytometer Reference Manual...
Page 144: ...144 BD FACSCanto II Flow Cytometer Reference Manual...
Page 152: ...152 BD FACSCanto II Flow Cytometer Reference Manual...