Empty the bleach-exposed waste.
5
Add approximately 1 L of bleach to the empty waste container (10 L
container).
6
If one month has passed since the last cap change:
•
Detach the cap from the trap.
•
Replace it with a new cap.
•
Write the date on the cap label.
7
Screw the cap assembly onto the tank and hand-tighten until it is fully
closed.
8
Re-attach the sensor and fluid lines.
To prevent tank overpressurization, do not overtighten the trap or attached
filter cap. Tighten each component only until it is hand-tight. Do not use
sealants such as Teflon® tape or other adhesives.
drain plug
trap
338677 Rev A
Waste (A)
space for date