Move the tube guide arm all the way forward.
The tube guide arm is needed only when using the Loader.
You are ready to load tubes automatically.
Maintaining the Loader
This section explains the cleaning procedure you should follow to keep your
Loader in good condition.
Spill Containment Area
The top of the Loader has a depressed area intended to contain spills that may
occur during operation. Check this area for fluids and clean as described in the
following section.
All instrument surfaces that come in contact with biological specimens can
transmit potentially fatal disease. Use universal precautions when cleaning
instrument surfaces. Wear suitable protective clothing and gloves.
To prevent damage, do not use isopropyl alcohol or ethanol on any Loader
surfaces.
automatic mode
manual mode