BD FACSCanto II Flow Cytometer Reference Manual
Purging the Bubble Filter
The bubble filter is situated between the plenum and the flow cell and ensures
that the flow cell remains bubble-free. However, should the fluidics run dry, you
might need to remove air from the bubble filter. Poor CVs might indicate a need
to purge the bubble filter.
1
Start the bubble filter purge:
•
In BD FACSCanto clinical software, select Cytometer > Cleaning
Modes > Bubble Filter Purge.
•
In BD FACSDiva software, select Instrument > Cleaning Modes >
Bubble Filter Purge and Degas Flow Cell.
2
Wait while the purge finishes.
3
Click OK when the completion message displays.
BD FACSCanto clinical software
BD FACSDiva software
BD FACSCanto clinical software
BD FACSDiva software