57
The DA-2006P can analyze pacemaker pulses as well as determine Refractory periods and
Sensitivity levels of on-demand pacemakers. For maximum versatility, the DA-2006P has
26 internally selectable pacemaker loads ranging from 50 ohms to 2300 ohms. The
DA-2006P can also test the noise immunity of the DUT by generating a 50 or 60 Hz noise
waveform with amplitude up to 100 mV. For sensitivity testing, the DA-2006P can output
a Square, Triangle or Haversine waveform with widths ranging from 10ms to 200ms. The
input circuitry of the DA-2006P is protected against damage in the case of an accidental
defibrillator discharge into the Pacemaker Input terminals.
The key is used to change to the Pacemaker Analyzer mode.
The Pacemaker Analyzer MAIN SCREEN shows the current CONFIGURATION, the TEST
RESULTS and the available FUNCTION KEYS.
The following is a sample of the PACE MAIN SCREEN:
NOTE: The Test Results section of the PACE MAIN SCREEN contains eight lines of data
that can be toggled to view the first 4 lines or the second 4 lines (See TEST RESULTS
section of manual for more information).
TRANSCUTANEOUS PACEMAKER ANALYZER
PACE MAIN SCREEN
Summary of Contents for DA-2006
Page 1: ...DEFIBRILLATOR ANALYZERS DA 2006 DA 2006P W PACER ANALYZER SERVICE MANUAL ...
Page 6: ...SECTION A USER MANUAL DA 2006 SERIES REV 06 ...
Page 7: ...DEFIBRILLATOR ANALYZERS DA 2006 DA 2006P W PACER ANALYZER USER MANUAL ...
Page 21: ...DA 2006 14 This page intentionally left blank ...
Page 37: ...Defibrillator Analyzer 30 This page intentionally left blank ...
Page 49: ...Defibrillator Analyzer 42 This page intentionally left blank ...
Page 55: ...Defibrillator Analyzer 48 This page intentionally left blank ...
Page 83: ...DA 2006 Series 76 This page intentionally left blank ...
Page 85: ...DA 2006 Series 78 This page intentionally left blank ...
Page 101: ...Auto Sequence Function 94 This page intentionally left blank ...
Page 113: ...Programming Auto Sequences 106 This page intentionally left blank ...
Page 115: ...DA 2006 Series 108 This page intentionally left blank ...
Page 122: ...115 NOTES ...
Page 123: ...SECTION A USER MANUAL DA CS 06 REV 02 ...
Page 124: ...DEFIBRILLATOR ANALYZER CALIBRATION STANDARD DA CS 06 USER MANUAL ...
Page 139: ...15 NOTES ...
Page 160: ...D5 ...
Page 162: ...D8 ...
Page 165: ...D12 ...
Page 169: ...D17 ...
Page 170: ...D18 ...
Page 171: ...NOTES ...