C 2
Use the following procedure to verify the Unit Under Test (UUT) operation:
DEFIBRILLATOR INPUT TEST
1. Set DMM to measure resistance
2. Measure the resistance between the UUT defibrillator input plates. Resistance
should be 49.5 to 50.5
Ω
.
3. Connect the Calibration Interface Module to the UUT. (See the DA-CS-06 User
Manual for more information.)
4. Set the UUT for High Range input mode.
5. Use Defibrillator to discharge at energy levels of 100, 200 and 360 Joules. Verify
that UUT measurements are within tolerance (2% of the calibrator measurement)
as shown on the PC display.
6. Run the UUT self-test waveform.
7. Measured power should be 100 to 150 Joules.
8. Set the UUT for Low Range Mode.
9. Use Defibrillator to discharge at energy levels of 5, 20, and 50 Joules. Verify that
UUT measurements are within tolerance (2% of the calibrator measurement) as
shown on the PC display.
NOTE: If Input Overload Warning, use next lower setting on source defibrillator.
10. Run the UUT self test waveform.
11. Measured Power should be 4 to 6 Joules
VERIFICATION TESTS
Summary of Contents for DA-2006
Page 1: ...DEFIBRILLATOR ANALYZERS DA 2006 DA 2006P W PACER ANALYZER SERVICE MANUAL ...
Page 6: ...SECTION A USER MANUAL DA 2006 SERIES REV 06 ...
Page 7: ...DEFIBRILLATOR ANALYZERS DA 2006 DA 2006P W PACER ANALYZER USER MANUAL ...
Page 21: ...DA 2006 14 This page intentionally left blank ...
Page 37: ...Defibrillator Analyzer 30 This page intentionally left blank ...
Page 49: ...Defibrillator Analyzer 42 This page intentionally left blank ...
Page 55: ...Defibrillator Analyzer 48 This page intentionally left blank ...
Page 83: ...DA 2006 Series 76 This page intentionally left blank ...
Page 85: ...DA 2006 Series 78 This page intentionally left blank ...
Page 101: ...Auto Sequence Function 94 This page intentionally left blank ...
Page 113: ...Programming Auto Sequences 106 This page intentionally left blank ...
Page 115: ...DA 2006 Series 108 This page intentionally left blank ...
Page 122: ...115 NOTES ...
Page 123: ...SECTION A USER MANUAL DA CS 06 REV 02 ...
Page 124: ...DEFIBRILLATOR ANALYZER CALIBRATION STANDARD DA CS 06 USER MANUAL ...
Page 139: ...15 NOTES ...
Page 160: ...D5 ...
Page 162: ...D8 ...
Page 165: ...D12 ...
Page 169: ...D17 ...
Page 170: ...D18 ...
Page 171: ...NOTES ...