Running a Defibrillator Test
49
A Cardioversion Test is simply an energy test with special attention being given to the
timing. The DA-2006 continuously monitors for the R-wave timing and displays, if
possible, the delay between the R-wave and the pulse. In Cardioversion testing, the
Defibrillator is set to deliver a pulse based on a specific delay after the R-wave.
(1) Connect ECG leads to the corresponding universal connector on the DA-2006.
The connectors are marked with both the AHA and International color codes.
(2) Turn on the DA-2006.
(3) The unit will come up in the “High Range Defibrillator” mode. This range is used
for normal adult testing.
NOTE: If it is desirable to run a test at 50 Joules or less with a peak
voltage of 1000 Volts or less, the unit may be changed to the “Low Range
Defibrillator” mode using the key.
CARDIOVERSION TEST
WARNING
This section is provided as a guide to familiarize
the user with the DA-2006 Series. It is not
intended to provide the necessary test sequence
for every Defibrillator. The user must consult the
manufacturer’s manual for each DUT to
determine the correct test procedure to follow.
Summary of Contents for DA-2006
Page 1: ...DEFIBRILLATOR ANALYZERS DA 2006 DA 2006P W PACER ANALYZER SERVICE MANUAL ...
Page 6: ...SECTION A USER MANUAL DA 2006 SERIES REV 06 ...
Page 7: ...DEFIBRILLATOR ANALYZERS DA 2006 DA 2006P W PACER ANALYZER USER MANUAL ...
Page 21: ...DA 2006 14 This page intentionally left blank ...
Page 37: ...Defibrillator Analyzer 30 This page intentionally left blank ...
Page 49: ...Defibrillator Analyzer 42 This page intentionally left blank ...
Page 55: ...Defibrillator Analyzer 48 This page intentionally left blank ...
Page 83: ...DA 2006 Series 76 This page intentionally left blank ...
Page 85: ...DA 2006 Series 78 This page intentionally left blank ...
Page 101: ...Auto Sequence Function 94 This page intentionally left blank ...
Page 113: ...Programming Auto Sequences 106 This page intentionally left blank ...
Page 115: ...DA 2006 Series 108 This page intentionally left blank ...
Page 122: ...115 NOTES ...
Page 123: ...SECTION A USER MANUAL DA CS 06 REV 02 ...
Page 124: ...DEFIBRILLATOR ANALYZER CALIBRATION STANDARD DA CS 06 USER MANUAL ...
Page 139: ...15 NOTES ...
Page 160: ...D5 ...
Page 162: ...D8 ...
Page 165: ...D12 ...
Page 169: ...D17 ...
Page 170: ...D18 ...
Page 171: ...NOTES ...