Running a Defibrillator Test
51
(8) The DA-2006 will automatically sense the voltage rise across the internal 50 ohm
load and begin taking readings. After the sampling is done (100 ms) the unit will
compute and display the results.
a. The power pulse is available at the oscilloscope output in real time at
200:1 signal attenuation when in low range and 1000:1 signal attenuation
when in high range.
b. After the computation, the pulse is automatically played back at a 200:1
time base expansion (200 times slower) on both the ECG leads and the
Paddle plates. The signal level is 1 mV per 1000 Volts on Lead 1.
c. At the same time, the test results are sent to the printer.
(9) The Status line will change to indicate the various steps as they are being done.
(10) At the end of the process the results are continuously displayed in the Test
Results section of the MAIN SCREEN. They will remain there until another test
is performed, the range is changed or the power is turned off.
NOTE: Special note should be made of the “Delay: xxx msec” line in the
results. This will show the delay between the peak of the R-wave and the
start of the Pulse.
Summary of Contents for DA-2006
Page 1: ...DEFIBRILLATOR ANALYZERS DA 2006 DA 2006P W PACER ANALYZER SERVICE MANUAL ...
Page 6: ...SECTION A USER MANUAL DA 2006 SERIES REV 06 ...
Page 7: ...DEFIBRILLATOR ANALYZERS DA 2006 DA 2006P W PACER ANALYZER USER MANUAL ...
Page 21: ...DA 2006 14 This page intentionally left blank ...
Page 37: ...Defibrillator Analyzer 30 This page intentionally left blank ...
Page 49: ...Defibrillator Analyzer 42 This page intentionally left blank ...
Page 55: ...Defibrillator Analyzer 48 This page intentionally left blank ...
Page 83: ...DA 2006 Series 76 This page intentionally left blank ...
Page 85: ...DA 2006 Series 78 This page intentionally left blank ...
Page 101: ...Auto Sequence Function 94 This page intentionally left blank ...
Page 113: ...Programming Auto Sequences 106 This page intentionally left blank ...
Page 115: ...DA 2006 Series 108 This page intentionally left blank ...
Page 122: ...115 NOTES ...
Page 123: ...SECTION A USER MANUAL DA CS 06 REV 02 ...
Page 124: ...DEFIBRILLATOR ANALYZER CALIBRATION STANDARD DA CS 06 USER MANUAL ...
Page 139: ...15 NOTES ...
Page 160: ...D5 ...
Page 162: ...D8 ...
Page 165: ...D12 ...
Page 169: ...D17 ...
Page 170: ...D18 ...
Page 171: ...NOTES ...