B 5
DEFIBRILLATOR CIRCUITRY
The defibrillator circuitry consists of a voltage input measurement section as well as an
analog output section. The input circuitry is used to measure the defibrillator discharge
pulse or the pacemaker pulse. The analog output is used to generate ECG or pacemaker
test waveforms.
The input circuit begins with the power resistors R22 and R25. These are 25
Ω
, 100 W,
Non-Inductive power resistors. These resistors absorb the power from the defibrillator
discharge pulse. Connected to these power resistors is a series of 16 low power resistors
R12 through R19 and R26 through R33. The purpose of these resistors is to divide the
voltage of the defibrillator discharge pulse. U1A is the differential amplifier that is used to
measure the divided voltage. The relays RL1 and RL2 are used along with R1, R2, R3, R6,
R7, and R8 to select the gain for either Low Range Defibrillator input, High Range
Defibrillator Input, or Pacemaker input measurements.
The amplifier U1:C is used to generate the zero offset voltage for the Defibrillator input
circuitry. U1:D is used to generate the self test waveform pulse. R4, R35, and RA1 are
used to properly attenuate the self-test and input signals and provide gain adjustment. The
output of U1:B is then connected to the A/D converter. The signal is offset by ½ of the A/D
reference voltage. This gives an analog range of 0 V = -2048 counts, 1.25V = 0 counts,
and 2.5 V = 2047 counts. The output of U1:A is also connected to the multiplexer U8 to
drive the scope output circuitry.
Summary of Contents for DA-2006
Page 1: ...DEFIBRILLATOR ANALYZERS DA 2006 DA 2006P W PACER ANALYZER SERVICE MANUAL ...
Page 6: ...SECTION A USER MANUAL DA 2006 SERIES REV 06 ...
Page 7: ...DEFIBRILLATOR ANALYZERS DA 2006 DA 2006P W PACER ANALYZER USER MANUAL ...
Page 21: ...DA 2006 14 This page intentionally left blank ...
Page 37: ...Defibrillator Analyzer 30 This page intentionally left blank ...
Page 49: ...Defibrillator Analyzer 42 This page intentionally left blank ...
Page 55: ...Defibrillator Analyzer 48 This page intentionally left blank ...
Page 83: ...DA 2006 Series 76 This page intentionally left blank ...
Page 85: ...DA 2006 Series 78 This page intentionally left blank ...
Page 101: ...Auto Sequence Function 94 This page intentionally left blank ...
Page 113: ...Programming Auto Sequences 106 This page intentionally left blank ...
Page 115: ...DA 2006 Series 108 This page intentionally left blank ...
Page 122: ...115 NOTES ...
Page 123: ...SECTION A USER MANUAL DA CS 06 REV 02 ...
Page 124: ...DEFIBRILLATOR ANALYZER CALIBRATION STANDARD DA CS 06 USER MANUAL ...
Page 139: ...15 NOTES ...
Page 160: ...D5 ...
Page 162: ...D8 ...
Page 165: ...D12 ...
Page 169: ...D17 ...
Page 170: ...D18 ...
Page 171: ...NOTES ...