C 5
VARIABLE RESISTANCE LOAD (DA-2006P ONLY)
1. Connect DMM to Variable Load Pacer inputs.
2. Set UUT Mode to Pace mode. Verify resistance of each Load selection
according to table below
Load
Selection
(ohms)
Valid DMM Measurement
of Resistance (ohms)
50
49.5 to 50.5
100
99 to 101
150
148.5 to 151.5
200
198 to 202
300
297 to 303
400
396 to 404
500
495 to 505
600
594 to 606
700
693 to 707
800
792 to 808
900
891 to 909
1000
990 to 1010
1100
1089 to 1111
1200
1188 to 1212
1300
1287 to 1313
1400
1379 to 1421
1500
1477.5 to 1522.5
1600
1576 to 1624
1700
1674.5 to 1725.5
1800
1773 to 1827
1900
1871.5 to 1928.5
2000
1970 to 2030
2100
2068.5 to 2131.5
2200
2167 to 2233
2300
2265.5 to 2334.5
Summary of Contents for DA-2006
Page 1: ...DEFIBRILLATOR ANALYZERS DA 2006 DA 2006P W PACER ANALYZER SERVICE MANUAL ...
Page 6: ...SECTION A USER MANUAL DA 2006 SERIES REV 06 ...
Page 7: ...DEFIBRILLATOR ANALYZERS DA 2006 DA 2006P W PACER ANALYZER USER MANUAL ...
Page 21: ...DA 2006 14 This page intentionally left blank ...
Page 37: ...Defibrillator Analyzer 30 This page intentionally left blank ...
Page 49: ...Defibrillator Analyzer 42 This page intentionally left blank ...
Page 55: ...Defibrillator Analyzer 48 This page intentionally left blank ...
Page 83: ...DA 2006 Series 76 This page intentionally left blank ...
Page 85: ...DA 2006 Series 78 This page intentionally left blank ...
Page 101: ...Auto Sequence Function 94 This page intentionally left blank ...
Page 113: ...Programming Auto Sequences 106 This page intentionally left blank ...
Page 115: ...DA 2006 Series 108 This page intentionally left blank ...
Page 122: ...115 NOTES ...
Page 123: ...SECTION A USER MANUAL DA CS 06 REV 02 ...
Page 124: ...DEFIBRILLATOR ANALYZER CALIBRATION STANDARD DA CS 06 USER MANUAL ...
Page 139: ...15 NOTES ...
Page 160: ...D5 ...
Page 162: ...D8 ...
Page 165: ...D12 ...
Page 169: ...D17 ...
Page 170: ...D18 ...
Page 171: ...NOTES ...