C 7
The Calibration procedure is used to adjust the UUT to comply with manufacturers
specifications. This procedure should only be performed if the UUT failed the
Defibrillator Input, Pacemaker Load Resistance or Pacemaker Input Verification tests.
After calibration is performed, the Verification test should be run to validate the new
calibration settings.
Part of the calibration can be performed Closed Box, other parts require Open Box. Closed
Box refers to the digital calibration adjustments that do not require the technician to open
the UUT.
NOTE: Opening the UUT will void the Manufacturer’s Warranty. Not all calibration
adjustments can be made Closed Box.
REQUIRED EQUIPMENT
The following equipment is needed for Validation / Calibration:
DMM, Fluke 45 or Equivalent
Battery Eliminator, 10VDC, 300mA, Regulated
Adjustable Power Supply, 0-12VDC, 500mA
Defibrillator (With pacemaker function for DA-2006P)
BC Biomedical Defibrillator Analyzer Calibration Interface Module (DA-CS-06)
PC Running the Calibration Interface Module software (DA-CS-06)
Oscilloscope, Lecroy 9304AM or equivalent
Centronics printer (DeskJet or Dot-Matrix)
CALIBRATION
Summary of Contents for DA-2006
Page 1: ...DEFIBRILLATOR ANALYZERS DA 2006 DA 2006P W PACER ANALYZER SERVICE MANUAL ...
Page 6: ...SECTION A USER MANUAL DA 2006 SERIES REV 06 ...
Page 7: ...DEFIBRILLATOR ANALYZERS DA 2006 DA 2006P W PACER ANALYZER USER MANUAL ...
Page 21: ...DA 2006 14 This page intentionally left blank ...
Page 37: ...Defibrillator Analyzer 30 This page intentionally left blank ...
Page 49: ...Defibrillator Analyzer 42 This page intentionally left blank ...
Page 55: ...Defibrillator Analyzer 48 This page intentionally left blank ...
Page 83: ...DA 2006 Series 76 This page intentionally left blank ...
Page 85: ...DA 2006 Series 78 This page intentionally left blank ...
Page 101: ...Auto Sequence Function 94 This page intentionally left blank ...
Page 113: ...Programming Auto Sequences 106 This page intentionally left blank ...
Page 115: ...DA 2006 Series 108 This page intentionally left blank ...
Page 122: ...115 NOTES ...
Page 123: ...SECTION A USER MANUAL DA CS 06 REV 02 ...
Page 124: ...DEFIBRILLATOR ANALYZER CALIBRATION STANDARD DA CS 06 USER MANUAL ...
Page 139: ...15 NOTES ...
Page 160: ...D5 ...
Page 162: ...D8 ...
Page 165: ...D12 ...
Page 169: ...D17 ...
Page 170: ...D18 ...
Page 171: ...NOTES ...