General Information
Page 24
Auto Syringe AS50 Infusion Pump Operator’s Manual
07-19-B3-430
Timed Infusion Modes
SINGLE DOSE:
In this mode, the pump is programmed to infuse a single
dose over a specified period of time. The dose is entered in
mL, and the time period is entered in hours and minutes.
MANUAL SCHEDULE:
In this mode, the pump is programmed to infuse a dose
over a specified period of time, and to repeat the dose at
specified intervals. The dose is entered in mL, and the time
period and interval are entered in hours and minutes.
The pump alerts the operator when a dose is due to start.
The operator must then press
START
to initiate each dose.
AUTO SCHEDULE:
In this mode, the pump is programmed to infuse a dose
over a specified period of time, and to repeat the infusion at
specified intervals. The dose is entered in mL, and the time
period and interval are entered in hours and minutes.
Whereas
MANUAL SCHEDULE
mode requires the operator
to press
START
to initiate each dose,
AUTO SCHEDULE
mode automatically starts each dose when it is due.