Simplified Instructions
Page 10
Auto Syringe AS50 Infusion Pump Operator’s Manual
07-19-B3-430
Note:
Steps 1 and 2 can be performed after programming the pump if necessary (for
example, if using CUSTOM DILUTION mode).
1.
Attach an IV extension set to the syringe and manually purge the air from the filled
syringe and tubing. Baxter recommends the following IV extension sets.
2.
Mount the syringe on the pump.
•
Pull out the plunger driver (2), slide it to top of the pump, and release it.
•
Using the lever (13) on the side or rear of the pump, release the barrel clamp (6).
•
Place the syringe in the cradle (7), making sure the syringe flange fits into
the slot (5).
•
Close the barrel clamp (6) firmly against the syringe barrel.
•
Pull out the plunger driver (2), slide it down to capture the top of the syringe
plunger, and push in firmly to capture syringe plunger.
3.
Slide the
ON/OFF
switch (14) to the
ON
position.
4.
Program the pump (see “Programming” on page 11).
5.
Purge the syringe and tubing again to remove slack from the plunger driver.
•
Press
PURGE
,
START
to begin purging. Repeat if necessary.
•
Press
STOP
if necessary to stop purging.
6.
Connect the IV extension set to the patient.
7.
Press
START
to begin the infusion.
Catalog
Number
Description
2N3345
36” Micro-Volume IV extension set
2N3346
36” High Flow Rate IV extension set
2N3347
36” Micro-Volume IV extension set with 0.22 micron filter
2N3348
60” Micro-Volume IV extension set
2N3349
60” High Flow Rate IV extension set
2N3350
60” Micro-Volume IV extension set with 0.22 micron filter
BARREL
FLANGE
PLUNGER