Detailed Instructions
Page 48
Auto Syringe AS50 Infusion Pump Operator’s Manual
07-19-B3-430
General Programming Information
Data Entry
The digit keys are used to enter program data.
Before pressing
CONFIRM
to complete a program data entry, the value can be
changed by pressing
▲
or
▼
or can be replaced by pressing
CLR
and entering
a new value. A value that appears automatically (as a result of selecting a
protocol) can be changed by pressing
▲
or
▼
or can be replaced by entering a
new value without pressing
CLR
.
Field Selection
The currently selected field on the pump’s status display flashes. Use
▲
and
▼
to select the next field. Press
EDIT
to open the currently selected field for
editing. When a field is opened for editing, the annunciators stop flashing and
the value itself begins flashing.
Editing in Standby State
All modes except CUSTOM DILUTION modes:
Almost all program data can be changed (edited). The exceptions are:
●
Once a protocol library, protocol or infusion mode has been selected, the
only way to select a different protocol library, protocol or infusion mode is
to turn the pump off, then turn it back on and reprogram.
●
The drug concentration (
CONC
.) and patient body weight (
BODY WT
) fields
cannot be changed (edited) once an infusion or bolus has been started.
To edit a programmed field:
1.
Use
▲
and
▼
to select the field to be changed.
2.
Press
EDIT
. Use
▲
,
▼
, and the digit keys as necessary to enter the new
value. Press
CONFIRM
to complete the change.