07-19-B3-430
Auto Syringe AS50 Infusion Pump Operator’s Manual
Page 45
Detailed Instructions
4.
If either the
INFUSE
field or the
OVER
field is edited, the pump treats the
result as a new infusion. As a reminder,
DOSE CANCELED
appears for a
few seconds.
5.
Pressing
CONFIRM
during delivery briefly displays additional program
data.
6.
Bolus operation is not available in this delivery mode.
Programming AUTO SCHEDULE Mode
AUTO SCHEDULE mode is a timed infusion mode for use when a
programmed dose is to be repeated at regular intervals. The infusion is
programmed in terms of dose volume and dose duration, and the time from
the start of one dose to the start of the next. Each dose delivery is
automatically initiated.
AUTO SCHEDULE mode also adds a delay to the start of the first dose,
allowing a time delay before the infusion regimen begins.
1.
At the
SELECT MODE
▲▼
prompt, use
▲
or
▼
to display the available
infusion mode selections. Press
CONFIRM
to select the desired infusion
mode.
2.
Use
▲
or
▼
and
CONFIRM
as required to select the syringe manufacturer
and size.
3.
Enter the dose volume (
INFUSE
) in mL. Press
CONFIRM
to complete the
data entry.
4.
Enter the dose duration (
OVER
) in hours and minutes. Press
CONFIRM
to
complete the data entry.
5.
Enter the interval (
EVERY
) in hours and minutes. This is the time from the
start of one dose to the start of the next dose. Press
CONFIRM
to complete
the data entry.
6.
Enter the delay (
NEXT DOSE IN
), then press
CONFIRM
. This is the time
interval from when
START
is pressed until the first dose becomes due.
Enter
0:00
to begin infusing the first dose immediately.
Example
: Program the pump to infuse 5 mL over 5 minutes, using a B-D 30 mL syringe.
Stop the infusion when the
TOTAL ML
field shows that about 2 mL have been delivered.
Now edit the
OVER
field to 6 minutes (note that
DOSE CANCELED
appears for a few seconds),
and start the infusion.
When the pump stops after 6 minutes, about 7 mL will have been delivered; 2 mL from the
original infusion, and 5 mL from the edited infusion.