07-19-B3-430
Auto Syringe AS50 Infusion Pump Operator’s Manual
Page 79
KEYPAD LOCKED
(single beep)
Pressing a key while the Lock
function is active.
Press
LOCK
to unlock the keypad.
LOAD SYRINGE
(single beep)
Attempt to start an infusion without
a syringe in place.
All three syringe detection sensors
have failed.
Mount a syringe after manually
purging.
If alert continues to occur, have the
pump serviced.
NO LIBRARY
(single beep)
VIEW LIBRARY
was selected and
no libraries exist.
Press
CONFIRM
to proceed to
VIEW MODES
prompt.
NO LIBRARY IS ENABLED
(single beep)
VIEW LIBRARY
was selected and
no libraries are enabled.
Press
CONFIRM
to proceed to
VIEW MODES
prompt.
NOT ALLOWED
(single beep)
The
BOLUS
key was pressed when
the bolus feature is not allowed by
the selected protocol.
Use the
NONE
protocol library
selection or the
NONE
protocol
selection (if available) if the bolus
feature is required.
OUT OF RANGE
(single beep)
No bolus is possible for the current
combination of syringe, body
weight, and drug concentration.
Use a different syringe or drug
concentration if the bolus feature is
required.
PUMP IS IDLE
(15 beeps and ALERT light flashes)
The pump has been in Standby
state for at least 2 minutes. Also
occurs if there is no keypress within
15 seconds after the infusion is
stopped.
The alert is self-cancelling. The
Lock function suppresses the audio
portion of this alert. Note: If the
keypad is locked, there will be no
beeps, but the
PUMP IS IDLE
message will still appear.
PUMP LIMIT
(single beep)
Attempting to program a value
outside the pump’s acceptable
range.
The pump displays the nearest
acceptable value. Press
CONFIRM
to accept displayed value, or enter a
different value and press
CONFIRM
.
SIZE MISMATCH
(single beep)
The detected syringe size is
different from the programmed size.
Check to be sure the correct
manufacturer has been
programmed. Press
CLR
if
necessary and reprogram the
syringe manufacturer and size.
SIZE UNKNOWN
(single beep)
The pump cannot identify the
syringe size for the selected
manufacturer.
Check to be sure the correct
manufacturer has been
programmed. Ensure that the barrel
clamp is fully closed. Press
CLR
if
necessary and reprogram the
syringe manufacturer and size.
USE DIGITS
(single beep)
Attempt to use
▲
or
▼
to initialize a
field containing dashes.
Use digit keys instead.
Message
Cause
Correction