Detailed Instructions
Page 34
Auto Syringe AS50 Infusion Pump Operator’s Manual
07-19-B3-430
Bolus in mL/hr and mL/min Modes
Note:
If a protocol has been selected, bolus delivery may be disallowed or a default
bolus size may appear automatically. If a default bolus size is displayed, use
▲
and
▼
or the digit keys to modify the value as necessary.
If the infusion program is to include a bolus delivery:
1.
Press
BOLUS
to obtain the bolus programming display.
2.
Press
EDIT
, then use the digit keys to set the bolus size (in mL).
3.
Press
CONFIRM
to store the bolus size and return to the normal display. The bolus
infusion rate is automatically set.
See “Bolus Function” on page 55 for more information on bolus delivery.
Note:
For mL/min mode, the bolus rate is displayed in mL/min. For all other modes, it is
displayed in mL/hr.
mL/hr and mL/min Programming Notes
1.
See “Syringe Data Tables” on page 25 for minimum and maximum infusion rates for
each syringe.
2.
Pressing
CONFIRM
during delivery briefly displays additional program data.
3.
A bolus can only be programmed, edited, or delivered after the syringe size has been
programmed.
Summary of Contents for Auto Syringe AS50
Page 1: ...Operator s Manual Auto Syringe AS50 Model AS50 Infusion Pump...
Page 99: ......