Additional Functions and Operations
Page 54
Auto Syringe AS50 Infusion Pump Operator’s Manual
07-19-B3-430
purge accumulated volume will disappear when the
START
key is pressed and will not
be included in the total volume.
4.
Repeat steps 2 and 3 as often as necessary to obtain a steady drip at the end of the
tubing.
Purge Delivery Table
Syringe
Manufacturer
Syringe Size
(mL)
Approximate
Purge
Volume (mL)
Purge Rate
(mL/hr)
Purge Rate
(mL/min)
B-D Plastic
1
3
5
10
20
30
60
0.02
0.08
0.15
0.22
0.39
0.50
0.75
11
38
74
107
188
241
366
0.19
0.63
1.2
1.7
3.1
4.0
6.1
Monoject
1
3
6
12
20
35
60
0.02
0.09
0.17
0.26
0.43
0.59
0.75
11
41
83
126
208
285
366
0.19
0.68
1.3
2.1
3.4
4.7
6.1
Terumo
1
3
5
10
20
30
60
0.02
0.09
0.18
0.27
0.44
0.58
0.75
11
41
87
129
210
276
438
0.19
0.69
1.4
2.1
3.5
4.6
7.3
NOTE:
Purge volume and rate information shown applies to pumps configured for Rate Range = H (High).
NOTE:
Purge volumes are approximately the amount delivered in a single complete purge delivery, without
slack in the drive system.
NOTE:
The volume of purged fluid is not added to the amount in the
TOTAL mL
field.