Additional Functions and Operations
Page 58
Auto Syringe AS50 Infusion Pump Operator’s Manual
07-19-B3-430
Bolus Editing
The bolus size can be edited during bolus programming and after a bolus has been fully
programmed. A bolus may be edited during an infusion without interrupting normal
delivery.
1.
Press
BOLUS
to display the
PRESS EDIT TO EDIT BOLUS SIZE
prompt.
2.
Press
EDIT
to open the bolus size field.
3.
Use
▲
and
▼
or the digit keys to modify the value as necessary, and press
CONFIRM
to complete the edit.
The pump displays
BOLUS READY
.
4.
Press
START
to begin bolus delivery immediately or press
CONFIRM
to save the bolus
data and return to the normal display.
Repeating a Bolus Delivery
To repeat a bolus delivery:
1.
Press
BOLUS
to display the bolus size.
2.
Review the bolus program data.
3.
Press
START
.
Changing the Syringe
To continue an infusion when a syringe becomes empty or to flush a line, the syringe may
be replaced when the pump is in the Standby state.
If the new syringe does not have the same barrel size as the currently programmed
syringe:
1.
The pump displays
VERIFY (NAME)
, where “name” refers to the currently programmed
syringe manufacturer.
Use
▲
and
▼
as needed to select the correct syringe manufacturer and press
CONFIRM
to complete the entry.
2.
The pump displays
VERIFY (SIZE)
where “size” is the new syringe size recognized by
the pump.
Press
CONFIRM
after verifying that the pump is displaying the correct syringe size.
Note:
Changing syringe size may invalidate the contents of other fields (e.g.
VOL LIMIT
).
In such cases, the pump automatically clears the necessary field(s). The infusion
cannot be continued until all cleared fields have been reprogrammed.