Auto Syringe AS50 Infusion Pump Operator’s Manual
Page 11
Simplified Instructions
Programming
Note:
If the Protocol Library feature is enabled, a
▲▼
SELECT LIBRARY
prompt is dis-
played first. Use
▲
,
▼
and
CONFIRM
as required to select a protocol library. Next,
select a protocol from the library.
If the selected protocol specifies a default infusion mode, the pump skips step 1.
1.
At the
SELECT MODE
▲▼
prompt, use
▲
or
▼
as necessary to display the available
infusion mode selections. Press
CONFIRM
to select an infusion mode.
2.
Select the syringe manufacturer and size using
▲
or
▼
as necessary and complete
the data entry by pressing
CONFIRM
. (For use of the Size Override feature, see “Size
Override” on page 64.)
3.
Enter the remaining program data as required. The current field is the one that is
flashing.
•
Press number keys to enter the desired value.
•
Press
CONFIRM
to complete the data entry.
4.
To change programmed data:
•
Press
▲
or
▼
to move to the desired field.
•
Press
EDIT
. The current value flashes.
•
Press
▲
,
▼
, or number keys to change the value.
•
Press
CONFIRM
to complete the data entry.
5.
To program or edit the bolus size:
•
Press
BOLUS
,
EDIT
.
•
Set the bolus size using the number keys.
•
Press
START
to deliver a bolus, or
CONFIRM
to store the bolus size.
6.
To clear the Total Delivered display:
•
Use
▲
or
▼
to move to the TOTAL mL field.
•
Press
EDIT
,
CLR
to reset the total to 0.00.