07-19-B3-430
Auto Syringe AS50 Infusion Pump Operator’s Manual
Page 15
General Information
Chapter 1
Technical Specifications
Model:
Auto Syringe AS50 Infusion Pump
Catalog Code:
1M8550
Size:
approx. 3.4" x 2.6" x 10" (8.6 x 6.7 x 25 cm)
Weight:
approx. 3.2 lb. (1.45 kg)
Flow Rate Accuracy:
±
3% of full scale plunger travel (not including syringe
tolerance)
Volume Accuracy:
±
3%, or
±
0.007" of travel, whichever is greater (not
including syringe tolerance)
Syringes:
B-D 1, 3, 5, 10, 20, 30, 60 mL plastic
Sherwood Monoject 1, 3, 6, 12, 20, 35, 60 mL
Terumo 1, 3, 5, 10, 20, 30, 60 mL
Flow Rate Range:
0.01 mL/hr to 438 mL/hr, depending on the syringe
Deliverable Volume:
Full syringe volume for 1-60 mL syringes
Data Display:
Self-prompting, multi-field LCD (Liquid Crystal Display)
Status Display:
Nine-LED (Light Emitting Diode) array
Power Requirement:
AC: 105-125V 60 Hz (battery charger)
DC: internal rechargeable battery pack
Battery Operating Time:
5 hours of operation at 100 mL/hr, or 12 hours of operation
at 2 mL/hr using a 60 mL syringe, following a charge of no
less than 16 hours.
Temperature Range:
50
o
F to 104
o
F (10
o
C to 40
o
C) Delivery of high viscosity
fluids at low temperatures is not recommended.
Keypad:
Elastomeric type, with tactile feedback
Construction:
Water-resistant, high-impact plastic case with removable
elastomeric protective bumpers