Maintenance
85
Operating Instructions HYGROPHIL® H 4230-10 Serie A BA 030520
Assembling the measuring chamber
Once parts have been cleaned or exchanged, the measuring chamber is
assembled again by reversing the steps taken to disassemble it.
Please observe the following instructions during assembly.
● The measuring chamber can only be assembled if the component parts
are in the right positions.
● When inserting the measuring cell into the middle section of the
measuring chamber, make sure that the pin on the measuring cell snaps
into place in the slot on the enclosure.
● Check all silicone hoses prior to reassembly. If they are no longer elastic
and dimensionally stable, they must be replaced. Replacement hoses
can be found among the accessories. Should you require more, you can
order replacement hose (see list of spare parts).
● Make sure that the air pressure measurement hoses are connected
correctly. The hoses must not be twisted or bent.
● Make sure that all O-rings are inserted correctly into the relevant slot and
that they don’t cant during assembly.
● Make sure that the water detector’s mount is located in the two enclosure
slots.